SIEMENS MAGNETOM SYMPHONY 1.5T

MODEL SIEMENS MAGNETOM SYMPHONY 1.5T
MODEL MAGNETOM Symphony
MAGNET Superconducting
POWER NEEDED LINE VOLTAGE, VAC 380/400/420/440/480
Kva (POWER NEEDED) 50
A/c, btu/hr (POWER NEEDED) 10.6 kW peak
CLINICAL USE (POWER NEEDED) Whole body
OPTIONS Whole-body integrated panoramic array, applications
packages: advanced cardiac, diffusion/perfusion, pMRI
BORE FEATURES (DIAMETER, cm) Lights, ventilation, nurse call, intercom
ECHO PLANAR IMAGING (CARDIAC GATING) Optional package
DICOM 3.0 COMPATIBLE (POWER NEEDED, VAC) Yes
SHIMMING (CRYOGEN USE, L/hr) Passive, active; 1st order; opt 2nd order
Diameter, cm (VACUUM) Body coil (DIAMETER, cm) 60
Head coil (DIAMETER, cm) 26 ID, open sides
ELECTROMAGNETIC FIELD STRENGTH, T (5-GAUSS FRINGE FIELD) 1.5
IMAGING
Pulse sequences (IMAGING)
GRE, IR, FIR, STIR, True IR/FISP, FSE, FLAIR, MT, SS-FSE, MT-SE, MTC, MSE, EPI, 3-D: DESS/CISS/ PSIF, GMR, fat/water sat/exc, others
Repetition time, msec (IMAGING) 2.4; opt 2/1.8, all @ 256 x 256
Echo time, msec (IMAGING) 1.1; opt 0.9/0.78, all @ 256 x 256
Inversion time, msec (IMAGING) 22
Slice thickness, mm (GANTRY) 0.1-200 (2-D), 0.05-20 (3-D)
Fov, cm (IMAGING) 0.5-50
Fov offsets (IMAGING) ±25
Scan orientations (IMAGING) Orthogonal (x, y, z); oblique second compound
Measuring matrix (IMAGING) 64 x 64 to 1024 x 1024
Display matrix (IMAGE DISPLAY) 1024 x 1024 full screen
Resolution, mm (PERFORMANCE) 0.021 in plane
Angiography Yes
IMAGING MODES Single (EYEPIECE) Yes
Multislice (IMAGING MODES) Yes
Volume study (IMAGING MODES) Yes
Additional (Alarms, high/low) Multi angle, multi oblique
SPECTROSCOPY (GRADIENT SUBSYSTEM) SVS or CSI optional
SURFACE COILS Spine (SURFACE COILS) Yes
Knee (SURFACE COILS) Yes
Neck (SURFACE COILS) Yes
Tmj (SYSTEM CAPABILITIES) Yes
Extremity (SURFACE COILS) Yes
Head (SURFACE COILS) Yes
Breast (APPLICATORS) Yes
Shoulder (SURFACE COILS) Yes
Additional (Alarms, high/low) Peripheral MRA, others
BORE SIZE, dia x L Or w x h x l, cm (BORE SIZE, dia x L) 60 dia
CARDIAC GATING Ecg/peripheral (CARDIAC GATING) Yes/Yes
Single slice, sec (RECONSTRUCTION TIME) 0.056@ full 256 x 256
Multislice, sec (RECONSTRUCTION TIME) 0.056/slice @ full 256 x 256
Volume, sec (RECONSTRUCTION TIME) 0.056/slice @ full 256 x 256
GRADIENT SUBSYSTEM Strength, mt/m (GRADIENT SUBSYSTEM) 2020/35; opt 30/52
Rise time to 10 mt/m, msec (GRADIENT SUBSYSTEM) 0.2; opt 0.13/0.08 1
Cpu (NETWORK/DATABASE) Dual Pentium IV
Memory size, mb (GRADIENT SUBSYSTEM) 72 GB HD, 2 GB RAM
ARRAY PROCESSOR (EXTERNAL BEAM) Dual Pentium IV
Memory size, mb (GRADIENT SUBSYSTEM) 5×18 GB HD, 1 GB RAM
Storage media/size (GRADIENT SUBSYSTEM) CD-ROM
Image storage capacity (GRADIENT SUBSYSTEM) Approx5,000
Respiratory gating (DISPLAY) Yes
Cooling method (X-RAY TUBE) Single cryogen, 2stage refrigeration
CRYOGEN USE, L/hr Liquid helium (CRYOGEN USE, L/hr) 0.075
MAGNET WEIGHT, kg (CRYOGEN USE, L/hr) 4,050; 5,500 in operation
DIMENSIONS (HXWXD),CM (CRYOGEN USE, L/hr) 235 x 215 x 160
5-GAUSS FRINGE FIELD Radial/axial, m (5-GAUSS FRINGE FIELD) 2.5/4
CONFIGURATION Compact, open sides
OTHER ATTRIBUTES (Interference compensation) 3-D shim; IPP/ panoramic table; 3-D MIP/MPR/SSD; remote diagnostics; Syngo/MRease; dynamic analysis; integrated post processing.
FDA CLEARANCE (Interference compensation) Yes
CE MARK (MDD) (Interference compensation) Yes

No Comments