Siemens Magnetom ‘C’ .3T MRI

MODEL MAGNETOM C!
CONFIGURATION OPTIONS
Range of magnetic field strength (CONFIGURATION OPTIONS) < 0.59 Tesla
Magnet Configuration (CONFIGURATION OPTIONS) Open bore – vertical/upright
Coil options (CONFIGURATION OPTIONS) Body, Head, Knee, Neck, Pelvic, Shoulder, Spine, Wrist, Flex Array, Phased Array
Gating Options (CONFIGURATION OPTIONS) Cardiac/ECG, Peripheral, Prospective with PACE, Respiratory
Are hardware/software modules available individually or are features available only as a system purchase (CONFIGURATION OPTIONS) Yes, modules can be purchased separately
TECHNICAL SPECIFICATIONS
Current release version (TECHNICAL SPECIFICATIONS) 2004 software
Gantry_Styles (TECHNICAL SPECIFICATIONS) Open_C Shaped
Magnet field strength (TECHNICAL SPECIFICATIONS) 0.35 Tesla
Gradient Strength (millitesla per meter) (TECHNICAL SPECIFICATIONS) 24 (effective 39) m T/M
FOV (Min, Max) (TECHNICAL SPECIFICATIONS) 0.5 – 40 cm
Reconstruction time – Single Slice (@ 256×256) (TECHNICAL SPECIFICATIONS) 355 recons per second (256square FFT, full FOV) 3226 recons per sec (256 square FFT, 25% recFOV) seconds
Reconstruction time – Multislice (40 slices @ 256×256) (TECHNICAL SPECIFICATIONS) 355 recons per second (256square FFT, full FOV) 3226 recons per sec (256 square FFT, 25% recFOV) seconds
Reconstruction time – Volume (@ 256×256) (TECHNICAL SPECIFICATIONS) 355 recons per second (256square FFT, full FOV) 3226 recons per sec (256 square FFT, 25% recFOV) seconds
Spin echo time (Specify matrix size) (TECHNICAL SPECIFICATIONS) min TE with SE 7.6ms (128 matrix ) 7.8ms (256 matrix) msec
Repetition time in msec (Specify matrix size) (TECHNICAL SPECIFICATIONS) 2.81 ms (128 matrix) 3.27 ms (256 matrix) msec
Inversion time in msec ( Specify matrix size ) (TECHNICAL SPECIFICATIONS) N/A msec
On-line image storage capacity (TECHNICAL SPECIFICATIONS) 110,000 (256 matrix) Images
Cryogen use (TECHNICAL SPECIFICATIONS) N/A with permanent magnet L/hour
Table weight limit [lbs] (TECHNICAL SPECIFICATIONS) 440 lbs.
Unit dimensions with service space (TECHNICAL SPECIFICATIONS) 137cm (W) + 209cm (D) + 189 (H)
Minimum room size requirements (Length 325 sq feet total install. Magnet room 16 x Width x Height) (TECHNICAL SPECIFICATIONS)
Power requirements (VAC, Amps)  Line voltage:380, 400, 420, 440, 480 V, 15 kVA
Heat output (TECHNICAL SPECIFICATIONS)  Cooling requirement: (BTU/ hr.) 15.6 kW BTU
Installed weight (TECHNICAL SPECIFICATIONS)  35 lbs ( 1 lbs = .45 Kg)
INTERFACE CAPABILITY AND SPECIFICATION
DICOM storage class user/provider  Yes
DICOM storage commit compliant (INTERFACE CAPABILITY AND SPECIFICATION)  Yes
DICOM modality worklist function (INTERFACE CAPABILITY AND SPECIFICATION)  Yes
DICOM print class user/provider (INTERFACE CAPABILITY AND SPECIFICATION)  Yes
DICOM Conformance URL (INTERFACE CAPABILITY AND SPECIFICATION)  www.siemensmedical.com
WARRANTY/SERVICE
Uptime Guarantee % (WARRANTY/SERVICE)  97% with Gold contract %
Availability of remote diagnostic services (WARRANTY/SERVICE)  Available
Technical assistance hotline number (WARRANTY/SERVICE)  1-800-767-2313
Warranty length for software [Month(s)] (WARRANTY/SERVICE)  12 Months
Warranty length on parts[Month(s)] (WARRANTY/SERVICE)  12 Months
Warranty length on labor [Month(s)] (WARRANTY/SERVICE)  12 Months
APPLICATIONS TRAINING
Quantity of applications training per installation (APPLICATIONS TRAINING)  Standard: 4 days off site syngo training for 2 tech. 4 days hands-on at clinical facility, 6 days on site training. Additional support available for fee per installation
Applications_training_locale (APPLICATIONS TRAINING)  Off-site, On-site
MARKET DATA/PRODUCT AVAILABILITY
Systems available/Qty. installed in US (MARKET DATA/PRODUCT AVAILABILITY) Aug. 2004; one installed and running
Systems available/Qty. installed in Europe (MARKET DATA/PRODUCT AVAILABILITY) July 2004; one installed and running
Systems available/Qty. installed in Asia (MARKET DATA/PRODUCT AVAILABILITY) June 2004; >5 installed and running
Systems available/Qty installed in ROW (MARKET DATA/PRODUCT AVAILABILITY) N/A
High and low range of unit/licenses price (APPLICATIONS TRAINING) N/A
Year product introduced (MARKET DATA/PRODUCT AVAILABILITY) 2004
MODEL SIEMENSE MAGNETOM C!
CONFIGURATION OPTIONS
RF frequency MHz (MAGNET SPECIFICATION) 14.6
Type (CONFIGURATION) Permanent
Shielding (MULTIPLE USE COMPONENTS) Passive
Homogeneity (5-GAUSS FRINGE FIELD) <4
40 cm DSV ppm (36cm DSV)
Number of measurement planes (MAGNET SPECIFICATION)  15
Number of points per plane (MAGNET SPECIFICATION) 24
Field stability ppm/hr (MAGNET SPECIFICATION) <±1.2/10min
COOLING SYSTEM (SYSTEMS) Heating system to stabilize permanent magnet
Patient aperture:
Width cm (Patient aperture at narrowest:) 137(side loading)
Depth cm (Patient aperture:) 137(side loading)
Height (pole to pole) cm (Patient aperture:) 42
Height(couch to pole) cm (Patient aperture:) 39
Magnet length with covers cm (Patient aperture:) 215
Scanner length cm 210 (table)
Patient couch:
Min. height cm (Patient couch:) 82
Max. height cm (Patient couch:) 82
Table top width cm (Patient couch:) 79
Body mass limit (full movement) kg (Patient couch:) 200
Mass: magnet only tonnes (MAGNET INSTALLATION DETAILS)  16
total tonnes (Patient couch:)  16.73
Depth (with covers) (y) cm (Patient couch:)  209
Width (with covers) (x) cm (MAGNET INSTALLATION DETAILS)  137
Height (with covers) (z) cm (Patient couch:)  189
Horizontal (x,y) 0.5 mT fringe field m (Patient couch:) 2.2×2.2
Vertical (z) 0.5 mT fringe field m (Patient couch:)  2.7
Minimum installed area m2* (MAGNET INSTALLATION DETAILS)  30
Minimum ceiling height cm* (MAGNET INSTALLATION DETAILS)  225
To include 0.5 mT fringe field
ELECTRONICS CABINETS
Number (STAINING POSTS)  1
Total width cm (ELECTRONICS CABINETS)  97
Maximum depth cm (ELECTRONICS CABINETS)  66
Maximum height cm (ELECTRONICS CABINETS)  196
COOLING SYSTEM (SYSTEMS) Air
Name/type/ version of the system (RF SYSTEM SPECIFICATION)
Number of independent RF receiver channels (standard/ optional) (RF SYSTEM SPECIFICATION)  2/4
Bandwidth of each independent RF receiver channel (MHz) (RF SYSTEM SPECIFICATION)  0.5
Number of Analog-to- Digital Converters for each independent RF channel (RF SYSTEM SPECIFICATION) 1 (reads out amplitude and phase)
Sampling frequency of each Analog-to-Digital Converter (MHz) (RF SYSTEM SPECIFICATION)  10
GRADIENT SYSTEM SPECIFICATION
Shielding (MULTIPLE USE COMPONENTS) Active
Single axis maximum amplitude
x mT/m (Single axis maximum amplitude) 24
y mT/m (Single axis maximum amplitude) 24
z mT/m (Single axis maximum amplitude) 24
Single axis slew rate (Single axis maximum amplitude)
x mT/m/ms (Single axis maximum amplitude) 55
y mT/m/ms (Single axis maximum amplitude) 55
z mT/m/ms (Single axis maximum amplitude) 55
Duty cycle at max amplitude (Single axis maximum amplitude) 1
Amplitude @100% duty cycle mT/m (RF SYSTEM SPECIFICATION) 24
Minimum TR ms (GRADIENT SYSTEM SPECIFICATION) 17
Spin echo*
Minimum TR ms (GRADIENT SYSTEM SPECIFICATION) 3.2
2D gradient echo*
Minimum TR ms (GRADIENT SYSTEM SPECIFICATION)  2.81
3D gradient echo*
Minimum echo ms spacing (3D gradient echo*)  3.76
Turbo spin echo*
Minimum echo ms spacing (3D gradient echo*)  0.72
Echo planar imagingâ€Â
Minimum TE ms (Echo planar imaging†)  82
Single-shot diffusion imaging with b-value of 1000 mm2/sâ€Â
* For 256×256 matrix
† For 128×128 matrix
RESOLUTION PARAMETERS
Minimum FOV mm (RESOLUTION PARAMETERS)  5
Maximum imaging matrix (RESOLUTION PARAMETERS)  512×512 1024×1024
Minimum 2D slice thickness mm (RESOLUTION PARAMETERS)  1.7
Minimum 3D slice thickness mm (RESOLUTION PARAMETERS)

No Comments